Ground Fault Interrupt (GFI)

Hipot and Dielectric Withstand test equipment can present an electrical shock hazard to the test operator. Test equipment manufacturers have included a GROUND FAULT INTERRUPT (GFI) circuit to help prevent the operator from being electrically shocked. A ground fault exists whenever currents i1 and i2 are not equal, as shown in the figure below.

When the current meter A1 and A2 detects the difference (i2-i1=iH) between the value i1 and actual i2 test current over high, the device cuts the power transiently to safely protect the equipment operator.

 

 