Hipot Test Equipment Comparison Charts

 

 

19052 Hipot Tester

 

Chroma19052

Instek GPI-735

Extech 7410
Associated Research 7650

Hioki 3159

Kikusui TOS8870A

Withstanding Voltage Test

Output voltage

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0.1-5KV
DC: 0.1-6KV

AC: 0-5KV
DC: 0-6KV

AC:
0-2.5 kV/
0-5kV

AC:
0-2.5 kV/
0-5kV

Load Regulation

±(1%+5V)

1%+5V

±(1% of output+5V)

±1.5% of reading

Better than 20%

Voltage Resolution

2V

5V

1V

Manual adjustment

NA

Voltage Accuracy

±(1%+5counts)

±(1%of reading + 10V)

±(2% of setting+5V)

±5% of reading

±1.5% of full Scale

Cutoff Current

AC: 30mA
DC: 10mA

AC: 0.1-40mA
DC: 0.1-10mA

AC: 30mA
DC: 10mA

AC: 0.01-120mA
DC: -

AC: 0.5/1/2/4/8/
10/100mA

Current Resolution

AC: 1uA
DC: 0.1uA

AC: 10uA
DC: 10uA

AC: 1uA
DC: 0.1uA

AC: 20uA
DC: -

-

Current Accuracy

±(1%+5counts)

±(1% reading +50uA)

±(2% f.S+2counts)

±3% of reading

±5% of high limit
±20% of low limit

Current Frequency

50Hz /60Hz

50Hz / 60Hz

50Hz /60Hz

SYNC AC Frequency

50Hz /60Hz

Test time

0.1-999sec,con

0.5-999.9sec

0,0.3-999.9sec

0.5-999sec

1-999sec

Ramp time

0.1-99.9sec,off

0.5-999.9sec

0.1-999.9sec

NA

 

ARC Detection Current

AC: 1-15/DC: 1-10mA

AC/DC: 1-40mA

AC/DC: 1-9Range

-

-

Insulation Resistance Test

IR Range

0.1M-50GW

1MW-10GW

1M-10GW

1M-2GW

1M-2GW

Output Voltage

50-1000V

50/100/500/1000V

100-1000V

500V/1000V

500V/1000V

Voltage Resolution

2V

4 Range

1V

(Only Two Range)

(Only Two Range)

Voltage Accuracy

±(1%+5V)

1%of reading +
10V

 

 

 500V:±5%
1000V:±10%

Resistance Accuracy

Typical:
±10%+10Count
10-50G:
±15%+10Count

1MΩ ~ 50MΩ:
±(5%Reading±1Count)
51MΩ ~ 200MΩ: ±(10%Reading±1Count)
201MΩ ~ 2000MΩ: ±(20%Reading±1Count)

1-1GW:±(2% of
Setting+2counts
1-10GW:±(8% of
Setting+2counts

±4% of
reading

 

 

OSC function
(patent)

YES

-

-

-

-

Back to top
Go back one page

19053 & 19035 Hipot Tester

 

Chroma 19053

Chroma 19035

Extech 7420
Associated Reasearch 7650

MicroTest 8820

Withstanding Voltage Test

Output voltage

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0-5KV
DC: 0-6KV

AC: 0-5KV
DC: 0-6KV

Load Regulation

±(1%+5V)

±(1% setting + 0.1%full Scale)

±(1% of output+5V)

-

Voltage Resolution

2V

2V

1V

10V

Voltage Accuracy

1%+5counts

1% setting + 0.1%full Scale

±(2% setting+
5V)

±5% setting

Cutoff Current

AC: 30mA
DC: 10mA

AC: 30mA
DC: 10mA

AC: 30mA
DC: 10mA

AC: 10mA
DC: 5mA

Current Resolution

AC: 1uA
DC: 0.1uA

AC: 1uA
DC: 0.1uA

AC: 1uA
DC: 0.1uA

-

Current Accuracy

±(1%+5counts)

±(1% Reading
+ 0.5% Range)

±(2% of setting+
2counts)

-

Current Frequency

50Hz /60Hz

50Hz /60Hz

50Hz /60Hz

50Hz /60Hz

Test time

0.1-999sec,continue

0.1-999sec, continue

0.1-999sec, continue

0.1-∞

Ramp up time

0.1-99.9sec,off

0.1-99.9sec,off

0.1-999.9sec

0.1-99.9sec

ARC Detection Current

AC: 1-15mA
DC: 1-10mA

AC: 1-15mA
DC: 1-10mA

1-9 Range

-

Insulation Resistance Test

Output Voltage

50-1000V

50-5000V

100-1000V

100-1000V

Voltage Resolution

2V

2V

1V

 

Voltage Accuracy

1.5%+5V

1%Set + 0.1%full Scale

±(2% of setting+
2counts)

±5%

IR Range

1M-10GW

1M-50GW

1M-10GW

1MΩ~10GΩ

Scanner Unit

8 ports, ±phase

8 ports, ±phase

8 ports, ±phase

8 ports, ±phase

OSC function
(patented)

YES

YES

-

-

DCR Measure

-

Y

-

-

Back to top
Go back one page

19071 Hipot Tester

 

CHROMA 19071

Instek 705

Extech 7130
Associated Research 3705

Hioki 3158

Kikusui TOS5050

Withstanding Voltage Test

Output Voltage

AC: 0.1-5KV

AC: 0.1-5KV

AC:0-5KV

AC:0-2.5KV
/0-5KV

AC:0-2.5KV
/0-5KV

Voltage Resolution

2V

5V

10V

Manual adjustment

10V

Voltage Accuracy

±(1%+5counts)

±3%of reading

±(2%of reading+2count)

±(1.5% of reading)

±1.5% of full Scale

Load Regulation

±(1%+5V)

±(1%+5V)

±(2% of output+5V)

?

±5% of full Scale

Cutoff Current

AC: 0.1-20mA

AC: 0.1-40mA

AC:0-20mA

AC: 0.01-120mA

AC: 0.1-110mA

Current Resolution

1uA

10uA

10uA

10uA

0.1-9.9mA:0.1maA
10-110mA:1mA

Current Accuracy

1.5%+5counts

±1% of reading

±(2% of reading+2count)

±3% of reading

±(5%+20uA) of
Upper cutoff

Current Frequency

50 / 60Hz

50 / 60Hz

50 / 60Hz

50 / 60Hz

50 / 60Hz

Test Time

0.3-999sec
continue

1-180sec

0.1-999sec

0.5-999sec

0.5-999sec

Ramp time

0.1-999sec,off

0-180sec,off

0,0.2-999,9sec

No

-

ARC
Detection Current

1-15mA

1-40mA

-

-

-

OSC function
(patented)

YES

-

-

-

-

Memory

10 Tests,
10 Steps Each

6 Tests,
6 Steps Each

10 Tests,
3 Steps Each

20 Tests

No

CE mark

Y

Y

N

N

Y

Back to top
Go back one page

19073 Hipot Tester

 

Chroma 19073

Instek 735

Extech 7142
Associated Research 3770

Hioki 3153

Kikusui TOS5051

Withstanding Voltage Test

Output voltage

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0.1-5KV
DC: 0.1-6KV

AC: 0-5KV
DC: 0-6KV

AC: 0.2-5KV
DC: 0.2-5KV

AC: 0-2.5/0-5KV
DC: 0-2.5/0-5KV

Load Regulation

±(1%+5V)

±(1%+5V)

±(1% of output+5V)

AC:7%
DC:16%

No Load
AC: 15%, DC: 3%

Voltage Resolution

2V

5V

10V

10V

10V

Voltage Accuracy

±(1%+5counts)

1%of reading + 10V

±(2%of reading+
5V)

±1.5% of reading
±2dgt

±1.5% of full
Scale

Cutoff Current

AC: 0.1-20mA
DC: 0.01-5mA

AC: 0.1-40mA
DC: 0.1-10mA

AC: 0-20mA
DC: 0-7.5mA

AC:0.01-100mA
DC:0.01-10mA

AC: 0.1-110mA
DC: 0.1-11mA

Current Resolution

AC: 1uA
DC: 0.1uA

AC: 10uA
DC: 10uA

AC: 1uA
DC: 0.1uA

AC: 10uA
DC: 10uA

AC: 10uA
DC: 10uA

Current Accuracy

1.5%+5Counts

1% reading +50uA

±(2% of setting+
2Counts)

±(2% f.S.+5 dgt)

±(5%+20uA) of
Upper cutoff

Current Frequency

50Hz / 60Hz

50Hz / 60Hz

50Hz / 60Hz

50Hz / 60Hz

50Hz / 60Hz

Test time

0.1-999sec,continue

 

0.1-999sec,con

0.3-999sec

0.5-999sec

Ramp time

0.1-999sec,off

 

0.1-999.9sec

0.1-99.9sec

NA

ARC Detection Current

AC: 1-15mA,DC: 1-5mA

AC/DC: 1-40mA

AC/DC: 0, 1-9 Range

-

-

Insulation Resistance Test

IR Range

0.1MW-50GW

1MW-10GW

1MW-10GW

0.1MW-10GW

-

Output Voltage

DC: 50-1000V

DC: 50-1000V

DC: 30-1000V

DC: 50-1000V

 

Voltage Resolution

2V

5V

10V

10V

-

Voltage Accuracy

5%+5Counts

1%of reading + 10V

±(2% of setting+2V)

±0.2 of reading

-

OSC function
(patent)

YES

-

-

-

-

Back to top
Go back one page

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19032 Safety Analyzer

 

Chroma 19032

Chroma 19032P

Instek 745A
Vitrek V4

Extech 7742
Associated Research 8106 / 7704

Withstanding Voltage Test

Output voltage

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0.05-5KV
DC: 0.05-6KV

AC: 0.1-5KV
DC: 0.1-6KV

AC/DC: 0-5KV

Load Regulation

1%+5V

1%+5V

1%+5V

±(2% of output+5V)

Voltage Resolution

2V

2V

5V

10V

Voltage Accuracy

±(1%+5counts)

±(1%+5counts)

±(1% of reading+10V)

±(2% of reading+2count)

Cutoff Current

AC: 40mA
DC: 12mA

AC: 100mA
DC: 20mA

AC: 40mA
DC: 10mA

AC: 0-40mA
DC: 0.3-20mA

Current Resolution

AC: 1uA
DC: 0.1uA

AC: 1uA
DC: 0.1uA

AC: 20uA
DC: 10uA

AC: 1uA
DC: 1uA

Current Accuracy

±(1%+5counts)

±(1%+5counts)

±(1% of reading+ 50uA)

±(2% of reading+2counts)

Current Frequency

50Hz / 60Hz

50Hz / 60Hz

50Hz / 60Hz

50Hz / 60Hz

Test time

0.3-999sec,con

0.3-999sec,con

1-180sec,con

0-999.9sec

Ramp up time

0.3-99.9sec,off

0.3-99.9sec,off

0-180sec,off

0.1-999.9sec,off

ARC Detection Current

AC: 1-40mA
DC: 1-12mA

AC: 1-40mA
DC: 1-12mA

40 Level

1-9 Range

Insulation Resistance Test

Output Voltage

0.05-1KV

0.05-1KV

50/100/500/1000V

50V-1000V

Voltage Resolution

2V

2V

5V

1V

Voltage Accuracy

±(2%+0.5% Range)

±(2%+0.5% Range)

±(2%+5V)

±(7% of reading+2count)

IR Range

1M-50GW

1M-50GW

1M-10GW

0.05M-50GW

Resistance Resolution

0.1MW

0.1MW

1MW

AVG 0.1MW

Resistance Accuracy

5% typical

5% typical

1-500MW: 5% of reading
501M-2GW: 10% of reading
2001M-10GW: 20% of reading

AVG
(7% of reading+2counts)

Ground Resistance Test

Output Current

AC: 1-30A 60A(option)

3-40A

DC: 3-42A

AC: 1-40A

Current Resolution

10mA

10mA

10mA

10mA

Current Accuracy

±(1%Read+0.2%Range)

±(1%Read+0.2%Range)

±1% of read±50mA

AVG ±(2.5% of reading+2counts)

GR Range

10-510mW

10-510mW

0-600mW

0-600mW

Resistance Resolution

0.1mW

0.1mW

100uW

1mW

Resistance Accuracy

±(1%Reading+0.1%Range)

±(1%Reading+0.1%Range)

±(1% of reading+2mW)

±(3% of reading+2counts)

Twin-Port Function

YES

YES

-

-

Scanner Unit (option)

8 ports, ±phase
5HV3GB , 3HV5GB
8HV,4HV4GB

8 ports, ±phase
5HV3GB , 3HV5GB
8HV,4HV4GB

8HV+8GB
16HV

8HV+8GB

Lc Probe

300V,20A,
RMS, Peak,P1-P2

300V,20A,
RMS, Peak,P1-P2

-

277V,15A
RMS

 

19032 Leakage Current Test Comparison

 

Chroma

Extech

Leakage Current

Model

6000-04

6000-07

6000-08

7740Opt.726

7742Opt.739

DUT Voltage
(Line input)

300V

300V

300V

277V

277V

DUT Voltage Accuracy

±(1.5%Reading
+0.2% full scale)

±(1.5%Reading
+0.2% full scale)

±(1.5%Reading
+0.2% full scale)

±(1.5%Reading
+2Counts)

±(1.5%Reading
+2Counts)

DUT Current

10A

20A

20A

15A

15A

Current Accuracy

±(1.5%Reading
+0.5% full scale)

±(1.5%Reading
+0.5% full scale)

±(1.5%Reading
+0.5% full scale)

±(2%Reading
+2Counts)

±(2%Reading
+2Counts)

Leakage Current
Measurement

RMS

RMS

RMS+ Peak
AC+DC

RMS

RMS+ Peak
AC+DC

Leakage Current Accuracy(RMS)

±(2%Reading
+1% full scale)

±(2%Reading
+1% full scale)

±(2%Reading
+1% full scale)

±(1.5%Reading
+3Counts)

±(1.5%Reading
+3Counts)

Leakage Current Accuracy(Peak)

-

-

±(5%Reading
+1% full scale)

-

±(10%Reading
+2uA)

P2 to G ,
P1 to P2

-

Y

Y

-

Y

Back to top
Go back one page